Cryogenic chamber
CR-2017 M1

 

Cr-2011 model M1 cryogenic chamber is a mobile chamber, dedicated for 1-2 people

 

It was developed based on experience from the construction of the Cr-2011 CH-1/B stationary chamber, and the container chamber. It is a leading achievement of CREATOR’s engineering thought.

Basic parameters of the cryochamber

  • Nominal current consumption 16 (A)
  • Electrical supply 230/50 (V/Hz)
    Temperature range in the treatment chamber -60 to -160oC
  • Temperature range in the atrium: from -40oCto -60oC
  • Recommended temperature in the treatment chamber -120oC
  • Recommended temperature in the atrium -60oC
  • Cooling time of the cryochamber to temp. -120oCto 30 – 40 (min)
  • Dwell time in the treatment chamber (adjustable by operator) 1-3 (min)
  • Dwell time in the vestibule (adjustable by operator) 1-3 (min)
  • Automatic cryochamber regeneration: user programmable time
  • Liquid nitrogen
  • Cooling of the cryochamber to the recommended temperature in the treatment chamber -120oC100 -120 (kg)
  • During the treatment (3 min) 3 – 4 (kg)
  • Consumption of liquid nitrogen during treatments – 80 kg/h
  • Power supply: alternating voltage
  • Atmospheric pressure: 96 – 106 kPa
  • Ambient temperature: 10 – 25oC
  • Relative humidity: < 40%
  • Type of work: continuous operation
  • Voltage value and frequency: 230V, 50 Hz
  • Rated power 1100 watts
  • Temperature in the atrium: -60oC+5oC
  • Maximum number of patients per chamber (depending on size): up to 1-2 patients
  • Operating temperature range of the treatment chamber: -60oCto -160oCin 1oCincrements.
  • Treatment time programmable in the range: up to 3 minutes in 1s increments
  • Oxygen concentration in the chamber: 18,5 – 21 %
  • Temperature stabilization in the treatment chamber: +/- 3 (max. +/-10)˚C
  • Amount of breathing air entering the chamber: (163 K) > 35 m3/h
  • Life span of oxygen sensors: 3-4 years to -160˚C
Decrease in oxygen concentration in the chamber:
– Below 17% – audible signal and automatic shut-off of cryogenic liquid supply to both the chamber and the building
– Above 22% – audible signal and automatic cut-off of cryogenic liquid supply to both the chamber and the building
For the patient: a button in the chamber – a light and sound signal.

General scheme
and layout of rooms