Cryogenic chamber CR-2017 M1
Cryogenic chamber Cr-2011 model M1 is a mobile chamber, dedicated for 1-2 people.
It was created on the basis of experience from the construction of stationary chamber Cr-2011 CH-1/B, and container chamber. It is a leading achievement of the engineering thought of CREATOR.
Basic operating parametrs
- Operating data:
- Medium and consumption
- Working conditions
- Rated working conditions
- Cryo-chamber operating parameters
- Alarm signaling
- Nominal current consumption 16 (A)
- Power supply 230/50 (V/Hz)
- Temperature range in the treatment chamber -60°C to -160°C
- Atrium temperature range: -40°C to -60°C
- Recommended temperature in the treatment chamber -120°C
- Recommended atrial temperature -60°C
- Cryochamber cooling time to -120°C is 30 - 40 (min).
- Duration of stay in the treatment chamber (adjustable by the operator) 1-3 (min)
- Atrium residence time (adjustable by the operator) 1-3 (min)
- Automatic regeneration of cryogenic chambers: user-programmable time
- Liquid nitrogen
- Refrigeration of the cryogenic chamber to the temperature recommended in the treatment chamber -120°C, 100 -120 (kg)
- Nitrogen consumption for cooling: up to -120°C -100 kg,
- Consumption of liquid nitrogen during the treatment (3 min) 35-80 (l)
- Consumption of liquid nitrogen during treatments - 35-70 l/h
- Power supply: alternating voltage
- Atmospheric pressure: 96 - 106 kPa
- Ambient temperature: 10 - 25°C
- Relative humidity: <40%
- Type of work: continuous work
- Voltage and frequency: 230 V, 50 Hz
- Rated Power 450 VA
- Power supply: AC voltage
- Atmospheric pressure: 96 - 106 kPa
- Ambient temperature: 10-25°C
- Relative humidity: < 40%
- Type of work: continuous work
- Voltage and frequency: 230V, 50 Hz
- Rated power 1100 W
- Temperature in the atrium: -60°C to -20°C
- Maximum number of patients in the chamber (depending on size): up to 1-2 people
- Working temperature range of the treatment chamber: -100°C to -160°C every 1°C
- Treatment time programmable in the range: up to 5 minutes every 1s
- Oxygen concentration in the chamber: 18,5 - 21 %
- Recommended temperature in the treatment chamber -120°C +/- 3 (max. +/-10) C
- The amount of breathing air entering the chamber: (163 K) > 35 m3/h
- Life span of oxygen sensors: 3-4 years to -160°C
- Decrease in oxygen concentration in the chamber:
- Less than 17% audible signal and automatic shut-off of cryogenic fluid supply to both the chamber and the building.
- Above 22% audible signal and automatic shut-off of cryogenic fluid supply to both the chamber and the building. - For the Patient: button in the chamber - light and sound signal.